Main Features:
Applied for adult and child
With high-intelligent control system
With LCD display for ventilation parameters
Comes with lightweight for transport
Main specification:
Power: 110 to 220V AC , 50 to 60Hz, built-in battery for backup
Gas source: O2
Ventilation mode: IPPV, SIMV, SIGH, SPONT and manual (optional: CPAP, PEEP)
Parameter monitoring: Tidal volume, minute volume, frequency, inhaling time, airway pressure and PEEP (optional: O2 Concentration)
Waveform display: Pressure-time (optional: Flow-time)
Alarm: No tidal volume, minute volume too high (low), airway pressure too high (low), AC power failure, backup battery too low.
Application |
Adult and pediatric patients for respiratory failure during emergency or transportation in ambulance |
Gas source |
Air or Oxygen |
Ventilation mode
|
IPPV,SIMV,A/C,SIGh,MANUAL |
Tidal volume |
0~1500ml |
I:E |
3:1~1:5 |
Breathing rate |
4~99/min,4~65 under SIMV |
FiO2 |
48%~100% |
Pressure trigger
|
-2~0kpa |
Monitoring parameters |
Vt,Mv,Breathing rate,Airway pressure |
Alarms |
No Vt,MV too high(low),Airway pressure too high(low),AC power off,back-up battery below voltage, |
Power |
AC 220V/50Hz,built-in battery |
Basic configuration |
main unit,oxygen cylinder 3L,silicon mask/circuit and other accessories |
|
|